Keyboard shortcuts

Press or to navigate between chapters

Press S or / to search in the book

Press ? to show this help

Press Esc to hide this help

Drop casting TEM grid preparation

Caution

VERY ROUGH DRAFT, NOT AUTHORITATIVE. This is a personal journey written by @bobleesj (Sangjoon Bob Lee) during his first time drop casting. Photos and notes are his own; the actual experiments were done by Guoliang Hu and Caitlyn Obrero. Terminology, step ordering, and parameter values may be wrong or incomplete. The resulting grids have not yet been imaged to confirm the preparation worked. Treat this page as a starting point to refer back to, not as an SOP. A trained user must verify everything before relying on it.

This page documents how a TEM grid was prepared by drop casting a nanoparticle solution. The grid is first plasma cleaned to make the carbon film hydrophilic, then a small volume of solution is pipetted onto the grid and allowed to dry in air.

Acronyms:

  • easiGlow : PELCO easiGlow glow discharge / plasma cleaner
  • TEM : transmission electron microscope

Materials:

  • TEM grids (e.g., carbon-coated copper, lacey carbon, holey carbon)
  • Nanoparticle solution(s) in labeled vials
  • Micropipette (0.5 to 10 µL range) with matching tips
  • Anti-capillary tweezers
  • Glass petri dish lined with lens paper or weighing paper
  • Aluminum grid-holder plate for the easiGlow
  • Filter paper or lens paper to wick excess solvent

Location

Drop casting is done in the PELCO easiGlow bay in Nucleus Labs W024 (Cryogenic Electron Microscopy) inside the Stanford ChEM-H Building.

Stanford ChEM-H Building entrance Nucleus Labs W024 entrance sign

The easiGlow workstation sits next to the rotary pump and shares the bench with pipettes, tip boxes, and a gold-sample coupon. Solutions, tweezers, and grid boxes stay on the same bench.

PELCO easiGlow plasma cleaner workstation with rotary pump

Overview

PhaseProcedure
Part 1: Plasma clean the gridsLoad grids on holder, run glow discharge
Part 2: Drop cast the solutionPipette sample onto grid, wick, dry
Part 3: Store prepared gridsTransfer dried grids to storage box

TODO: Record approximate time for each phase once the workflow has been repeated a few times.

Part 1: Plasma clean the grids

Plasma cleaning removes organic contamination and makes the carbon support film hydrophilic, so the aqueous droplet spreads evenly instead of beading up. Skip this step and the solution will roll off the grid or concentrate at the edges.

1.1 Retrieve clean grids

  1. Open the grid box

    A user must keep grids in a labeled grid box between uses to avoid contamination. Open the box gently: the grids are thin enough that a gust of air can flip them.

    Opened blue grid box with a single grid visible in one slot
  2. Lift a grid with a fine-tip tool

    Use a wooden toothpick tip, carbon-tipped vacuum pen, or anti-capillary tweezers to remove a grid. Hold the grid by the edge: pressing on the center will punch a hole through the carbon film.

    Lifting a TEM grid from a blue grid box with a fine-tip pickup tool

1.2 Load grids onto the easiGlow plate

  1. Place grids shiny (carbon) side up on the aluminum plate

    The plate has rows of shallow circular wells sized for standard 3 mm grids. Shiny side up ensures the carbon film faces the plasma and becomes the hydrophilic surface.

    Two TEM grids placed shiny side up on aluminum grid-holder plate in a glass petri dish

    If you cannot tell which side is shiny, tilt the grid under the bench light. The side with the carbon film reflects more uniformly.

  2. Transfer the plate onto the easiGlow stage

    Slide the plate onto the stainless steel stage inside the easiGlow chamber. Center it under the upper electrode.

    Aluminum grid-holder plate placed on the easiGlow stage under the upper electrode

1.3 Run the glow discharge

  1. Close the chamber

    Lower the bell jar onto the O-ring. Press down gently and evenly until you feel the seal engage.

    Closing the easiGlow bell jar chamber onto the base O-ring
  2. Select the negative-glow program on the touchscreen

    On the easiGlow control panel, set the following parameters:

    ParameterValue
    ModeNEGATIVE
    GasGAS 1
    PressureTODO: confirm target pressure
    CurrentTODO: confirm current (mA)
    Glow time60 s
    Hold timeTODO: confirm hold time
    easiGlow touchscreen showing NEGATIVE mode, GAS 1, glow time 60 s

    TODO: Record the exact pressure, current, and hold-time values from the staff-defined recipe the next time the instrument is used.

  3. Start the cycle

    Press Auto. The pump pulls the chamber down to the set pressure, then the plasma ignites (a faint purple glow appears inside the bell jar). The cycle auto-vents at the end.

  4. Use the grids promptly

    Plasma-cleaned grids lose hydrophilicity over time as the surface re-adsorbs ambient hydrocarbons. Prepare your solution and pipette tips before starting the glow discharge so you can drop cast immediately.

    TODO: Confirm how long after glow discharge the grids remain usable.

Part 2: Drop cast the solution

2.1 Prepare the solution

  1. Check that your sample vials are labeled and in a rack

    Each vial should have a unique label (e.g., S1-3, S2-1) matching your lab notebook. Vortex or gently invert any solution that has sat for more than a few minutes to resuspend settled particles.

    Rack of labeled sample solution vials on the bench
  2. Mount a fresh tip on the micropipette

    Use a tip that matches your pipette (e.g., SL-10 XLS, 0.5 to 10 µL). A fresh tip prevents cross-contamination between samples.

    Mounting a fresh tip onto the micropipette from a tip rack

2.2 Pipette onto the grid

  1. Draw solution from the vial

    Press the plunger to the first stop, dip the tip into the solution, and release slowly to draw the volume. Avoid drawing air: a bubble at the tip will spray the droplet off the grid.

    TODO: Record the volume actually used (microliters) and the pipette model/range.

    Drawing solution from an eppendorf tube with a micropipette
  2. Deposit the droplet on the grid

    Hold the pipette vertically above the grid. Touch the droplet to the grid surface so it wets across the film. Do not press the tip into the carbon film.

    Depositing colored nanoparticle solutions onto grids on the aluminum plate

    If the droplet beads up and rolls off, the grid was not plasma cleaned recently enough. Repeat Part 1 with a fresh grid.

2.3 Wick and dry

  1. Transfer the grid onto a piece of lens paper

    Place the grid sample-side up on clean lens paper or filter paper. The paper wicks excess solvent from the edges without disturbing the deposited particles.

    Drop-cast grids drying on lens paper with visible dark spots of deposited solution
  2. Let the grid air dry

    Wait until the droplet is fully evaporated. Drying time depends on the solvent and droplet volume.

    Do not blow on the grid to speed drying. Breath moisture deposits contamination, and forced airflow can redistribute particles asymmetrically.

    TODO: Record actual drying time for the solvent(s) used.

Part 3: Store prepared grids

  1. Pick up the dried grid with anti-capillary tweezers

    Grip the grid by the edge only. The dark dot visible in the center is the dried sample: touching it will scrape particles off.

    Picking up a dried drop-cast grid from lens paper with anti-capillary tweezers
  2. Return the grid to a labeled grid box

    Note the well number and sample ID in your lab notebook.

    Aluminum grid storage plate with numbered wells

Troubleshooting

TODO: This table is generic drop casting guidance, not validated against the actual SNSF workflow. Verify with Guoliang Hu or Caitlyn Obrero before relying on it.

SymptomLikely causeAction
Droplet beads up and rolls offGrid not plasma cleaned or surface re-contaminatedRe-run glow discharge with a fresh grid
Particles concentrated at droplet edge (coffee-ring)Solvent evaporated too slowly or grid not levelReduce droplet volume, or wick the edge with filter paper
Carbon film tornPipette tip touched the gridLower the droplet without contacting the film
No plasma ignitionChamber not at base pressureReseat the bell jar, wipe the O-ring
Visible aggregates on gridSolution not resuspendedVortex sample, dilute, repeat

Changelog

  • Apr 22, 2026 - Initial draft captured during first drop casting session. Photos and notes by @bobleesj. Experiments by Guoliang Hu and Caitlyn Obrero. Output not yet verified by TEM imaging.